                              | Reasons for empty pipe alarm by  electromagnetic flow meter during use |  30-04-2020 
                           
                              
                                Whether the measuring fluid  fills the measuring tube.Check whether the signal cable connection  of the electromagnetic flow meter is correct.Check whether the sensor  electrode is normalMake the flow rate of the  electromagnetic flow meter zero, and the observation shows that the  conductivity ratio should be less than 100%.In the case of current, the  resistance of the measuring terminals SIGI1,SIG2 and SIGGND should be less than  50 k¦¸Use a multimeter to measure the DC voltage between DS1  and DS2 should be less than1V,otherwise ,it means that the sensor electrode is  contaminated and should be cleaned. 
 